EasyManua.ls Logo

Minebea Intec PR5900 - Page 122

Default Icon
382 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
[Max at]
Perform calibration with weights, without weight (by mV/V values or data), or
linearization, see Chapter 4.7.1.2.1.4.
- [by load], see Chapter 4.7.1.2.1.4.1.
- [by mV/V], see Chapter 4.7.1.2.1.4.2.
- [by data] Load cell data (smart calibration), see Chapter 4.7.1.2.1.4.3.
- [Linear.] (linearization), see Chapter 4.7.1.2.1.4.4.
WP-A
Max 3000g d= 0.1g
00003
TST
+
Calibration @admin
1
2
3
Max
Scale interval
Dead load at
Max at
calibrated at
Sensitivity
3000.0 g
0.1 g
mV/V
mV/V
mV/V
µV/d
0.040750
1.177246
0.987513
0.470898
30000 d
1 d
102.832 g
3000.000 g
2516.5 g
98.11 cnt/d
by load by mV/V by data Linear. Save
Calculate the reference value
for the weighing electronics test...
This window appears briey.
The maximum capacity (Max) is displayed with the ID "TST" and without a weight unit.
The conrmation is displayed as follows: "saving calibration…"
Note:
After exiting calibration, go to the menu [Param] and set the parameter [Settings
locked] to activate overwrite protection via the software (see Chapter 2.7.2).
Note:
A unique check number is created every time a calibration or changed parameters
are saved. This can be viewed in the menu [Operating] - [System information] -
[Show calibration check numbers] (see Chapter 5.7) and may also be written on the
W&M label.
To save an application for use in legal metrology, see Chapter 4.7.1.2.
4.7.1.2.3 Dead load at (correction)
In this menu item, the dead load (empty scale or vessel) is redened for an existing
calibration of the Internal weighing point weighing electronics.
If the vessel/platform weight changes by an amount that is higher than the zero range;
e.g. due to dead load reduction, dead load increase, or mechanical changes, the functions
for automatic zerotrack and manual zero setting no longer work.
the range currently used by zerotrack or set zero is displayed in the menu [Operating]-
[System information]- [HW options]- [WPx]- [Monitor] .
3. Press the [Save] soft key to save the calibration.
Process Controller Maxxis 5 PR 5900 4 System setup
Minebea Intec EN-120

Table of Contents