Press the key to switch on increased resolution (10-fold) for the weight value.
If the full zeroset range is already being utilized, you can still correct the dead load
(overwrite protection must be deactivated, see Chapter 2.7) without aecting the other
calibration data/parameters.
To this end, calibration is performed in the menu [Operating] - [System setup] -
[Weighing points] - [Weighing point x] - [Modify] - [Dead load at] - [By load] (see
Chapter 4.7.1.2.1.3).
Note:
If a linearization was carried out (see Chapter 4.7.1.2.1.4.4), the following message
appears once the parameter has been selected: "Cannot be changed while linearization
is active."
Only deleting of the linearization points deactivates the linearization mode!
4.7.1.3 Units
In this menu item, additional weight units for the display and the display accuracy of the
Internal weighing point weighing electronics are congured.
The device is calibrated with a selected weight unit. Up to two more units can also be
displayed.
The weight display is switched using the function keys (F1 or F2). This must be assigned
beforehand, see Chapter 4.3.
When three units are displayed, they are shown in the following sequence: from 1 to 2, 3,
1, 2, and so on.
If unit 2 or 3 is displayed, the symbol (for non-legal weights) is also shown in
addition to the weight unit on devices used for legal metrology.
In the following example,
- unit 2 is dened in [oz]. The display accuracy is set to [basic accuracy].
This means that the scale interval in weight roughly corresponds to the calibrated
scale interval.
- Unit 3 in [lb] has a display accuracy that is one level higher, meaning that the scale
interval is roughly one level smaller than the calibrated scale interval.
The values for Max and Min for the additional units are (roughly) calculated taking the
associated scale intervals into account.
Accessible via MENU - [Operating] - [System setup] - [Weighing points] - [Weighing
point x] - [Units] .