For range selection for scales and settings, refer to Chapter 4.7.1.1.2 and 4.7.1.1.3
[Default] soft key
Settings are reset to factory settings.
4.7.7.3 Calib
In this menu item, calibration is performed for the digital weighbridge load cell Pendeo
®
Truck.
- [Quick install with search load cells and set dead load], see Chapter 4.7.7.3.1.
- [Search for connected load cells], see Chapter 4.7.7.3.2.
- [View and assign load cells], see Chapter 4.7.7.3.3.
- [Calibrate the scale] (load cells), see Chapter 4.7.7.3.4.
- [New], see Chapter 4.7.7.3.4.1.
- [Modify], see Chapter 4.7.7.3.4.2.
- [Assign load cell name], see Chapter 4.7.7.3.5.
- [Service function for load cells], see Chapter 4.7.7.3.6.
- [Corner correction], see Chapter 4.7.7.3.7.
During calibration, no data is changed in the digital load cells. The calibration data and
parameters are saved in the device. The unique serial numbers of the connected load cells
are monitored.
For the calibration the following order must be followed:
- Search for load cells and set dead load, see Chapter 4.7.7.3.1 or search for load cells
only, see Chapter 4.7.7.3.2.
- Assign load cells, see Chapter 4.7.7.3.3.
- Recalibrate: Maximum capacity with weight unit, scale interval, dead load, calibration
weight, see Chapter 4.7.7.3.4.1.
- Perform a corner correction if necessary; see Chapter 4.7.7.3.7.
Note:
For further information about calibrating weighing points, see Chapter 4.7.1.
Accessible via MENU - [Operating] - [System setup] - [Weighing points] - [Weighing
point x] - [Calib] .
3. Press the [Save] soft key to save the settings.