[Zerotrack]
The zero display is automatically maintained within set limits.
Selection: No/Yes
When [No] is selected, the next three parameters are not shown.
When [Yes] is selected, values for the next three parameters must be entered.
[Zerotrack range]
Range within which automatic zerotrack compensates for deviations.
Setting range: 0.25…10000.00 d
For use in legal metrology a value of 0.25 d must be entered.
[Zerotrack step]
If a weight change exceeds the adjusted value, automatic tracking does not function
any more.
Setting range for automatic tracking increments: 0.25…10 d
For use in legal metrology a value of 0.25 d must be entered.
WP-D
Max 200t d= 0.02t
0000
t
Weighing points @admin
1
2
3
Zerotrack time
Overload
Minimum weight
Range mode
1.0 s
9 d
20 d
single range
Default Save
[Zerotrack time]
Time interval for automatic zerotrack.
Setting range: 0.1…25 s
For use in legal metrology a value of 1 s must be entered.
[Overload]
Weighing range above the maximum capacity (Max) without error message.
Setting range: 0…9999999 d
For use in legal metrology a value of max. 9 d = e must be set.
[Min. weight]
Minimum weight at which a print command can be triggered.
Setting range: 0…9999999 d
For use in legal metrology a value of at least 20 d must be set.
[Range mode]
Selection: single range, multi-range, multi-interval
Process Controller Maxxis 5 PR 5900 4 System setup
Minebea Intec EN-166