The average deviation of the individual load cells is calculated. Monitoring is
indicated via symbol above the weight unit.
Setting range: 0 to 100%.
[Standstill time]
The parameters [Standstill time] and [Standstill range] are used to dene the
standstill of the scale (stable balance).
The [Standstill time] parameter is entered in seconds. The permitted range is:
0.00…2 s.
The time can be entered from 0.00 to 2.00 seconds, but always makes up at least one
measurement time.
WP-D
Max 200t d= 0.02t
0000
t
Weighing points @admin
1
2
3
Standstill range
Tare timeout
Zeroset range
Zerotrack
Zerotrack range
Zerotrack step
1.00 d
2.5 s
50.00 d
Yes
0.25 d
0.25 d
Default Save
[Standstill range]
As long as the weight uctuations remain within this range, the device is determined
to be stable.
The [Standstill range] parameter is entered in "d". The permitted range is:
0.01…10.00 d.
For use in legal metrology, ≤1 d must be selected.
[Tare timeout]
Timeout for a tare/zeroset command that cannot be executed (e.g. due to mechanical
instability of the scale, incorrect lter setting, resolution too high, standstill condition
too strict).
This parameter is given in seconds. The permitted range is: 0.0…[2.5]…25 s.
At 0.0 s taring is only carried out when the scale is already stable.
[Zeroset range]
Determine a ±range around the zero point determined by the dead load during
calibration; within this range
- the displayed gross weight can be set to zero by pressing the zero-setting key (or
by a corresponding external command), and
- automatic zero tracking is active.
Setting range: 0.00…10000.00 d
For use in legal metrology a value ≤2% of the max must be entered, example: 60 d for
3000 e of Class III.