EasyManua.ls Logo

Minebea Intec PR5900 - Page 214

Default Icon
382 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11.000 t+
Corner correction @admin
LC 1
LC 2
LC 3
LC 4
6.317 t
4.854 t
2.421 t
5.102 t
Calc
You are free to choose any desired order.
11.000 t+
Corner correction @admin
LC 1
LC 2
LC 3
LC 4
4.321 t
3.287 t
4.133 t
6.960 t
Calc
When corner correction is completed, it is marked with [Correction ok].
WP-D
Max 200t d= 0.02t
Min 0.4t
0000
t
Calibration @admin
Calibration weight
Corner correction
11.000 t
correction ok
Vessel Save
4. Position (here: LC 1) is highlighted.
5. Conrm position.
6. Remove the calibration weight.
7. Repeat steps 1 to 6 for the remaining load cells.
8. Once all load cells have been loaded one time, press the [Calculate] soft key to
perform corner correction.
The total weight remains unchanged. Only the eect of the individual load cells is
corrected.
9. Press the [Save] soft key to save the corner correction.
Process Controller Maxxis 5 PR 5900 4 System setup
Minebea Intec EN-212

Table of Contents