Training Guide Course No. 196514
Level 2 Maintenance, X-1000 Series Dispensing Systems 4-148 P/N 196515 (Revision A)
Information Sheet 4-13-3
DP-3000 Troubleshooting (Continued)
I. Stopcock not actuating
Possible Cause Recovery
 Air not hooked up  Ensure that the air is on and is reaching the stopcock.
J. Inconsistent Accuracy And Repeatability
Possible Cause Recovery
 Lead screw malfunction  Contact Asymtek Technical Service Group for repair.
 Lead screw bearing
malfunction
 Contact Asymtek Technical Service Group for repair.
K. Needle Moves During Dispensing
Possible Cause Recovery
 Stopcock and needle
assembly is not assembled
correctly
 See “Preventing Needle Movement” in the DP-3000
Operations Manual.