4 Patient Alarms and INOPs
62
*** Brady (Pulse) SpO2, Press,
cmResp
Pulse rate < the extreme brady alarm limit. numeric flashes and alarm limit is
highlighted, red alarm lamp, alarm
tone
*** Bradypnea cmRR The respiration rate has dropped below the
bradypnea limit.
numeric flashes and alarm limit is
highlighted, red alarm lamp, alarm
tone
** CCO High
** CCI High
CCO Continuous Cardiac Output or CC Index is
above the high alarm limit.
numeric flashes and high alarm
limit is highlighted, yellow alarm
lamp, alarm tone
** CCO Low
** CCI Low
CCO Continuous Cardiac Output or CC Index is
below the low alarm limit.
numeric flashes and low alarm limit
is highlighted, yellow alarm lamp,
alarm tone
** cmRR High cmResp The respiration rate has exceeded the high
alarm limit.
numeric flashes and high alarm
limit is highlighted, yellow alarm
lamp, alarm tone
** cmRR Low cmResp The respiration rate has dropped below the low
alarm limit.
numeric flashes and low alarm limit
is highlighted, yellow alarm lamp,
alarm tone
** CPP High CPP The CPP value has exceeded the high alarm
limit.
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
** CPP Low CPP The CPP value has fallen below the low alarm
limit.
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
*** Desat
*** Desat xx < yy
SpO
2
The SpO
2
value has fallen below the
desaturation alarm limit. xx denotes the lowest
measured value, and yy is the desaturation limit.
numeric flashes, red alarm lamp,
alarm tone
** EC10 Alarm
*** EC10 Alarm
at Information Center
EC10 IntelliBridge A yellow (**) or red (***) patient alarm is
present on the IntelliBridge module. Check the
monitor display for more detailed alarm
information.
(on monitor) yellow or red alarm
lamp; the alarm text is defined by
the IntelliBridge device driver
* End AFIB ECG/Arrhythmia Atrial fibrillation no longer detected for the
configured Afib end delay time
yellow alarm lamp, short yellow
alarm tone
* End Irregular HR ECG/Arrhythmia Irregular HR no longer detected for the
configured irregular HR end delay time.
yellow alarm lamp, short yellow
alarm tone
** et<Agent Label> High GM The end tidal agent high alarm limit has been
exceeded.
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
** et<Agent Label> Low GM The end tidal agent value has fallen below the
low alarm limit
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
** etCO High CO
2
, AGM, GM The end tidal CO
2
high alarm limit has been
exceeded.
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
** etCO Low CO
2
, AGM, GM The end tidal CO
2
value has fallen below the
low alarm limit.
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
** etO High O
2
, AGM The end tidal O
2
high alarm limit has been
exceeded.
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
Alarm Message From Condition Indication