4 Patient Alarms and INOPs
63
** etO Low O
2
, AGM The end tidal O
2
value has fallen below the low
alarm limit.
numeric flashes, and low limit is
highlighted, yellow alarm lamp,
alarm tone
* Event
** Event
*** Event
at Information Center
Event surveillance An event has occurred and the event
notification is configured to alarm. Check on
the monitor for more details on event group.
(on monitor) event group name
flashes, yellow or red alarm lamp
and alarm tone
* Event:<Event Group>
** Event:<Event Group>
*** Event:<Event Group>
Event surveillance An event has occurred and the event
notification is configured to alarm.
event group name flashes, yellow or
red alarm lamp and alarm tone
*** Extreme Brady
ECG Heart rate < the extreme brady alarm limit. numeric flashes and alarm limit is
highlighted, red alarm lamp, alarm
tone
*** Extreme Tachy ECG Heart rate > the extreme tachy alarm limit. numeric flashes and alarm limit is
highlighted, red alarm lamp, alarm
tone
* HR High
** HR High
ECG Heart rate > the high HR alarm limit numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone. If configured to short
yellow, the sound switches off after
5 seconds if Arrhythmia is switched
on.
* HR Low
** HR Low
ECG Heart rate < the low HR alarm limit numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone. If configured to short
yellow, the sound switches off after
5 seconds if Arrhythmia is switched
on.
** imCO High GM The inspired minimum CO
2
high alarm limit
has been exceeded
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone.
** in<Agent Label> High GM The inspired agent high alarm limit has been
exceeded
numeric flashes, high limit is
highlighted, yellow alarm lamp,
alarm tone
** in<Agent Label> Low GM The inspired agent value has fallen below the
AGT low alarm limit
numeric flashes, low limit is
highlighted, yellow alarm lamp,
alarm tone
** inNO High GM The inspired N
2
O high alarm limit has been
exceeded.
numeric flashes, high limit is
highlighted, yellow alarm lamp,
alarm tone.
** inO High GM The inspired O
2
high alarm limit has been
exceeded.
numeric flashes, high limit is
highlighted, yellow alarm lamp,
alarm tone.
** inO Low GM The inspired O
2
value has fallen below the low
alarm limit
numeric flashes, low limit is
highlighted, yellow alarm lamp,
alarm tone
***inO Low Oxygen GM The inspired O
2
value has fallen below 18
vol%.
numeric flashes, low limit is
highlighted, red alarm lamp, alarm
tone
** IPI Low CO
2
The IPI value has fallen below the low alarm
limit.
numeric flashes, and low limit is
highlighted, yellow alarm lamp,
alarm tone
Alarm Message From Condition Indication