4 Patient Alarms and INOPs
82
cl NBP Batt Low
INOP tone
The remaining battery time of the NBP Pod is below 2 hours.
cl NBP Batt Malf
INOP tone
There is a malfunction in the NBP Pod's battery system. Contact your service personnel.
cl NBP Batt Temp
INOP tone
The temperature of the battery in the NBP Pod is critically high. Check that the Pod is not covered or
exposed to a heat source. If the INOP persists, remove the Pod from the patient and contact your
service personnel.
cl NBP Check Batt
INOP tone
The battery in the NBP Pod is nearing the end of its useful life. Only 50 charge/discharge cycles
remain. Contact your service personnel to replace the battery.
cl NBP Chk Chrg IF The interface between the NBP Pod and the Charging Station is not working properly. Try cleaning the
contacts and reposition the Pod on the charging station.
cl NBP ChkSettings The NBP Pod's settings may have changed or become corrupt. Check the settings.
cl NBP Disconnect
INOP tone
The NBP Pod has lost the connection to the monitor.
cl NBP License Req The NBP Pod requires a valid software license for operating.
cl NBP No Cradle
INOP tone
The NBP Pod is not in its cradle.
cl NBP Remove
INOP tone
The temperature of the battery in the NBP Pod is too high. Remove the Cableless Measurement
Device from the patient and contact service personnel.
cl NBP Serv Batt
INOP tone
The battery in the NBP Pod has reached the end of its useful life. It can no longer be charged. Contact
your service personnel to replace the battery.
cl Resp Batt Empty
!!cl Resp BatEmpty
!!!cl Resp BtEmpty
Corresponding INOP tone
The remaining battery time of the Respiration Pod is below 30 minutes. Charge the battery.
cl Resp Batt Incmp
INOP tone
The battery in use with the Respiration Pod is incompatible. Replace it with one approved for use with
the Respiration Pod.
cl Resp Batt Low The remaining battery time of the Respiration Pod is below 2 hours.
cl Resp Batt Malf There is a malfunction in the Respiration Pod's battery system. Contact your service personnel.
cl Resp Batt Temp The temperature of the battery in the Respiration Pod is critically high. Check that the Pod is not
covered or exposed to a heat source. If the INOP persists, remove the Pod from the patient and
contact your service personnel.
cl Resp Check Batt The battery in the Respiration Pod is nearing the end of its useful life. Only 50 charge/discharge cycles
remain. Contact your service personnel to replace the battery.
cl Resp Chk ChrgIF The interface between the Respiration Pod and the Charging Station is not working properly. Try
cleaning the contacts and reposition the Pod on the charging station.
cl Resp Chk Sett The Respiration Pod's settings may have changed or become corrupt. Check the settings.
cl Resp Disconnect The Respiration Pod has lost the connection to the monitor.
cl Resp Remove The temperature of the battery in the Respiration Pod is too high. Remove the Pod from the patient
and contact service personnel.
cl Resp Serv Batt The battery in the Respiration Pod has reached the end of its useful life. It can no longer be charged.
Contact your service personnel to replace the battery.
cl SpO₂ Batt Empty
!!cl SpO₂ BatEmpty
!!!cl SpO₂ BtEmpty
Corresponding INOP tone
The remaining battery time of the SpO
2
Pod is below 30 minutes. Charge the battery.
cl SpO₂ Batt Incmp
INOP tone
The battery in use with the SpO
2
Pod is incompatible. Replace it with one approved for use with the
SpO
2
Pod.
INOP Message, Indication What to do