4 Patient Alarms and INOPs
83
cl SpO Batt Low
INOP tone
The remaining battery time of the SpO
2
Pod is below 2 hours.
cl SpO Batt Malf
INOP tone
There is a malfunction in the SpO
2
Pod's battery system. Contact your service personnel.
cl SpO Batt Temp
INOP tone
The temperature of the battery in the SpO
2
Pod is critically high. Check that the Pod is not covered or
exposed to a heat source. If the INOP persists, remove the Pod from the patient and contact your
service personnel.
cl SpO Check Batt
INOP tone
The battery in the SpO
2
Pod is nearing the end of its useful life. Only 50 charge/discharge cycles
remain. Contact your service personnel to replace the battery.
cl SpO Chk ChrgIF The interface between the SpO
2
Pod and the Charging Station is not working properly. Try cleaning
the contacts and reposition the Pod on the charging station.
cl SpO Chk Sett The SpO
2
Pod's settings may have changed or become corrupt. Check the settings.
cl SpO Disconnect
INOP tone
The SpO
2
Pod has lost the connection to the monitor.
cl SpO No Cradle
INOP tone
The SpO
2
Pod is not in its cradle.
cl SpO Remove
INOP tone
The temperature of the battery in the SpO
2
Pod is too high. Remove the SpO
2
Pod from the patient
and contact service personnel.
cl SpO Serv Batt
INOP tone
The battery in the SpO
2
Pod has reached the end of its useful life. It can no longer be charged. Contact
your service personnel to replace the battery.
clResp License Req The Respiration Pod requires a valid software license for operating.
clSpO License Req The SpO
2
Pod requires a valid software license for operating.
cmResp NeoPatient? The Respiration Pod cannot be used with neonatal patients. Check the patient category.
cmResp No Signal No signal is being detected from the Respiration Pod. Check that the Pod is attached correctly with the
adhesive attachment and that it moves with the patient's breathing.
cmRespCan'tAnalyze The Respiration Pod algorithm cannot reliably analyze the signal. Respiration rate and pulse rate are
affected
cmRespPediPatient? The Respiration Pod cannot be used with pediatric patients. Check the patient category.
INOP Message, Indication What to do