Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1030User Manual 1178.8520.02 ─ 01
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:HOLDoff <Duration>
This command defines for how long the the signal at the marker position is demodula-
ted.
In the time domain continuous demodulation is always on.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<Duration> Range: 10 ms to 1000 s
*RST: Marker demodulation = OFF
Default unit: S
Example:
CALC:MARK:FUNC:DEM:HOLD 3s
Manual operation: See " Marker Stop Time " on page 458
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:SELect <DemodMode>
This command selects the demodulation mode at the marker position.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<DemodMode> AM
AM demodulation
FM
FM demodulation
*RST: AM
Example:
CALC:MARK:FUNC:DEM:SEL FM
Manual operation: See "Type" on page 354
See " Modulation " on page 458
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation[:STATe] <State>
This command turns the audio demodulator on and off when the measurement rea-
ches a marker position.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
Analyzing Measurements (Basics)