Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1100User Manual 1178.8520.02 ─ 01
[SENSe:]CORRection:TRANsducer:SCALing <ScalingType>
This command selects the frequency scaling of the transducer factor.
Parameters:
<ScalingType> LINear | LOGarithmic
*RST: LINear
Manual operation: See " X-Axis Scaling " on page 598
[SENSe:]CORRection:TRANsducer:SELect <Name>
This command selects a transducer factor.
Parameters:
<Name> String containing the name of the transducer factor.
If the name does not exist yet, the R&S FSV/A creates a trans-
ducer factor by that name.
Example:
CORR:TRAN:SEL 'FACTOR1'
Manual operation: See " Activating / Deactivating " on page 596
See " Create New Line " on page 596
See " Name " on page 598
[SENSe:]CORRection:TRANsducer[:STATe] <State>
This command turns the selected transducer factor on or off.
Before you can use the command, you have to select a transducer.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Manual operation: See " Activating / Deactivating " on page 596
[SENSe:]CORRection:TRANsducer:UNIT <Unit>
This command selects the unit of the transducer factor.
Before you can use the command, you have to select and turn on a transducer.
Parameters:
<Unit> string as defined in table below
*RST: DB
Example:
CORR:TRAN:UNIT 'DBUV'
Manual operation: See " Unit " on page 598
String Unit
'DB' dB
'DBM' dBm
Configuring the R&S FSV/A