Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1127User Manual 1178.8520.02 ─ 01
Example: Define the system language:
SYST:LANG '8563E'
Set the response back to factory setting:
SYS:REV:FACT
Query the revision:
REV?
Response:
920528
Usage: Event
Manual operation: See "Resetting the Factory Revision" on page 705
SYSTem:SHIMmediate ONCE
Executes any received remote commands that cause changes to the hardware and
have not been executed yet due to a SYST:SHIM:STAT OFF command.
Example:
SYST:SHIM:STAT ON
...
SYST:SHIM ONCE
Usage: Event
Manual operation: See "Set Hardware Immediately" on page 702
SYSTem:SHIMmediate:STATe <State>
Determines when the remote commands that change hardware settings on the
R&S FSV/A are executed.
Regardless of this setting, the firmware automatically sets the hardware when a sweep
is started.
This setting is not changed by the preset function.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Remote commands that cause changes to the hardware are
only executed when the SYSTem:SHIMmediate ONCE com-
mand is executed.
ON | 1
Remote commands are always executed immediately when they
are received by the instrument.
*RST: 1
Example:
SYST:SHIM:STAT ON
...
SYST:SHIM ONCE
Manual operation: See "Set Hardware Immediately" on page 702