INSTrument:COUPle:CENTer <State>
This command turns synchronization of the frequency between measurement channels
on and off.
Parameters:
<State> ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: ALL
Example:
INST:COUP:CENT ALL
Synchronizes the center frequency.
INSTrument:COUPle:DEMod
<State>
This command turns synchronization of the audio demodulator configuration between
measurement channels on and off.
Parameters:
<State> ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: NONE
Example:
INST:COUP:DEM ALL
Synchronizes the audio demodulator configuration.
INSTrument:COUPle:GAIN <State>
This command turns synchronization of the preamplifier configuration between mea-
surement channels on and off.
Parameters:
<State> ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: NONE
Example:
INST:COUP:GAIN ALL
Synchronizes the preamplifier configuration.
Configuring the R&S FSV/A