EasyManua.ls Logo

R&S FSV3000 Series - Page 1139

R&S FSV3000 Series
1268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1139User Manual 1178.8520.02 ─ 01
INSTrument:COUPle:CENTer <State>
This command turns synchronization of the frequency between measurement channels
on and off.
Parameters:
<State> ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: ALL
Example:
INST:COUP:CENT ALL
Synchronizes the center frequency.
INSTrument:COUPle:DEMod
<State>
This command turns synchronization of the audio demodulator configuration between
measurement channels on and off.
Parameters:
<State> ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: NONE
Example:
INST:COUP:DEM ALL
Synchronizes the audio demodulator configuration.
INSTrument:COUPle:GAIN <State>
This command turns synchronization of the preamplifier configuration between mea-
surement channels on and off.
Parameters:
<State> ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: NONE
Example:
INST:COUP:GAIN ALL
Synchronizes the preamplifier configuration.