Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
246User Manual 1178.8520.02 ─ 01
The following tables show the child nodes and attributes of each element and show if a
child node or attribute is mandatory for the R&S FSV/A to interpret the file or not. The
hierarchy of the XML cannot be seen in the tables. View one of the predefined files
already stored on the R&S FSV/A in the "C:\Program
Files\Rohde-Schwarz\FSV3000\<version>\sem_std" directory, or check the
structure as shown below.
Below, a basic example of the structure of the file is shown, containing all mandatory
attributes and child nodes. Note that the PowerClass element and the Range element
are themselves elements of the BaseFormat element. They must be inserted where
noted. They are separated here simply to provide a better overview. Also, no example
values are given here to allow a quick reference to the tables above. Italic font shows
the placeholders for the values.
●
The BaseFormat element is structured as follows:
– <RS_SEM_ACP_FileFormat Version="1.0.0.0">
<Name>"Standard"</Name>
<Instrument>
<Type>"Instrument Type"</Type>
<Application>"Application"</Application>
</Instrument>
<LinkDirection Name="Name">
<ReferencePower>
<Method>"Method"</Method>
</ReferencePower>
<PowerClass Index="n">
<!-- For contents of the PowerClass node, see Table 7-14 -->
<!-- Define up to four PowerClass nodes -->
</PowerClass>
</LinkDirection>
</RS_SEM_ACP_File>
●
The "PowerClass" element is structured as follows:
– <PowerClass Index="n">
<StartPower Unit="dBm" InclusiveFlag="true" Value="StartPowerValue"/>
<StopPower Unit="dBm" InclusiveFlag="false" Value="StopPowerValue"/>
<DefaultLimitFailMode>"Limit Fail Mode"</DefaultLimitFailMode>
<Range Index="n">
<!-- For contents of the Range node, see Table 7-15 -->
<!-- Define up to twenty Range nodes -->
</Range>
…
</PowerClass>
●
The "Range" element is structured as follows:
– <Range Index="n">
<Name="Name">
<ChannelType>"Channel Type"</Channel Type>
<WeightingFilter>
<Type>"FilterType"</Type>
<RollOffFactor>"Factor"</RollOffFactor>
Spectrum Emission Mask (SEM) Measurement