Common Measurement Settings
R&S
®
FSVA3000/ R&S
®
FSV3000
386User Manual 1178.8520.02 ─ 01
Remote command:
[SENSe:]BANDwidth[:RESolution]:TYPE on page 893
Default Coupling
Sets all coupled functions to the default state ( "Auto" ). In addition, the ratio "RBW/
VBW" is set to "Sine[1/1]" and the ratio "Span/RBW" to 100.
For more information see Chapter 8.5.1.3, "Coupling VBW and RBW", on page 378.
Remote command:
[SENSe:]BANDwidth[:RESolution]:AUTO on page 892
[SENSe:]BANDwidth:VIDeo:AUTO on page 893
[SENSe:]SWEep:TIME:AUTO on page 897
Sweep/Average Count
Defines the number of sweeps to be performed in the single sweep mode. Values from
0 to 200000 are allowed. If the values 0 or 1 are set, one sweep is performed.
The sweep count is applied to all the traces in all diagrams.
If the trace modes "Average" , "Max Hold" or "Min Hold" are set, this value also deter-
mines the number of averaging or maximum search procedures.
In continuous sweep mode, if "Sweep Count" = 0 (default), averaging is performed
over 10 sweeps. For "Sweep Count" =1, no averaging, maxhold or minhold operations
are performed.
For more information, see Chapter 8.5.1.8, "How Much Data is Measured: Sweep
Points and Sweep Count", on page 381.
Remote command:
[SENSe:]SWEep:COUNt on page 895
Sweep Points
Defines the number of measured values to be collected during one sweep.
For details see Chapter 8.5.1.8, "How Much Data is Measured: Sweep Points and
Sweep Count", on page 381.
Remote command:
[SENSe:]SWEep[:WINDow<n>]:POINts on page 896
Optimization
In FFT mode, several FFT analysis steps are required to cover the entire measurement
span. The span which is covered by one FFT analysis step is called partial span. The
partial span cannot be defined directly, but it can be optimized according to measure-
ment requirements.