Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
897User Manual 1178.8520.02 ─ 01
[SENSe:]SWEep:TIME <Time>
This command defines the sweep time. It automatically decouples the time from any
other settings.
Parameters:
<Time> refer to data sheet
*RST: depends on current settings (determined automati-
cally)
Default unit: S
Manual operation: See " Sweep Time " on page 163
See "Sweep Time" on page 396
[SENSe:]SWEep:TIME:AUTO
<State>
This command couples and decouples the sweep time to the span and the resolution
and video bandwidths.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
SWE:TIME:AUTO ON
Activates automatic sweep time.
Manual operation: See " Harmonic Sweep Time " on page 290
See " Sweep Time " on page 384
See " Default Coupling " on page 386
[SENSe:]SWEep:TYPE <Type>
This command selects the sweep type.
Parameters:
<Type> AUTO
Automatic selection of the sweep type between sweep mode
and FFT.
FFT
FFT mode
*RST: AUTO
Example:
SWE:TYPE FFT
Manual operation: See " Sweep Type " on page 387