Common Measurement Settings
R&S
®
FSVA3000/ R&S
®
FSV3000
411User Manual 1178.8520.02 ─ 01
Changing the Automatic Measurement Time ( Meastime Manual )
This function allows you to change the measurement duration for automatic setting
adjustments. Enter the value in seconds.
Note: The maximum possible measurement duration depends on the currently
selected measurement and the installed (optional) hardware. Thus, the measurement
duration actually used to determine the automatic settings may be shorter than the
value you define here.
Remote command:
[SENSe:]ADJust:CONFigure:DURation:MODE on page 920
[SENSe:]ADJust:CONFigure:DURation on page 919
Upper Level Hysteresis
When the reference level is adjusted automatically using the Auto Level function, the
internal attenuators and the preamplifier are also adjusted. To avoid frequent adapta-
tion due to small changes in the input signal, you can define a hysteresis. This setting
defines an upper threshold the signal must exceed (compared to the last measure-
ment) before the reference level is adapted automatically.
Remote command:
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer on page 920
Lower Level Hysteresis
When the reference level is adjusted automatically using the Auto Level function, the
internal attenuators and the preamplifier are also adjusted. To avoid frequent adapta-
tion due to small changes in the input signal, you can define a hysteresis. This setting
defines a lower threshold the signal must fall below (compared to the last measure-
ment) before the reference level is adapted automatically.
Remote command:
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer on page 920