Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
920User Manual 1178.8520.02 ─ 01
[SENSe:]ADJust:CONFigure:DURation:MODE <Mode>
In order to determine the ideal reference level, the R&S FSV/A performs a measure-
ment on the current input data. This command selects the way the R&S FSV/A deter-
mines the length of the measurement .
Parameters:
<Mode> AUTO
The R&S FSV/A determines the measurement length automati-
cally according to the current input data.
MANual
The R&S FSV/A uses the measurement length defined by
[SENSe:]ADJust:CONFigure:DURation on page 919.
*RST: AUTO
Manual operation: See " Resetting the Automatic Measurement Time ( Meastime
Auto )" on page 410
See " Changing the Automatic Measurement Time ( Meastime
Manual )" on page 411
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer <Threshold>
Parameters:
<Threshold> Range: 0 dB to 200 dB
*RST: +1 dB
Default unit: dB
Example:
SENS:ADJ:CONF:HYST:LOW 2
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level falls below 18 dBm.
Manual operation: See " Lower Level Hysteresis " on page 411
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer <Threshold>
Parameters:
<Threshold> Range: 0 dB to 200 dB
*RST: +1 dB
Default unit: dB
Example:
SENS:ADJ:CONF:HYST:UPP 2
Example: For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level rises above 22 dBm.
Manual operation: See " Upper Level Hysteresis " on page 411
[SENSe:]ADJust:CONFigure:TRIGger <State>
Defines the behavior of the measurement when adjusting a setting automatically (using
SENS:ADJ:LEV ON, for example).
Setting Basic Measurement Parameters