Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
753User Manual 1178.8520.02 ─ 01
Parameters:
<Standard> For more information see Chapter 7.3.9, "Reference: Predefined
CP/ACLR Standards", on page 193.
If you want to load a customized configuration, the parameter is
a string containing the file name.
Manual operation: See " Predefined Standards " on page 159
See " User Standards " on page 160
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:CATalog?
This command queries all files containing ACLR standards.
Suffix:
<n>
.
Window
<m> Marker
<sb> Multi-SEM: 1 to 3
for all other measurements: irrelevant
Return values:
<Standards> List of standard files.
Usage: Query only
Manual operation: See " User Standards " on page 160
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:DELete
<Standard>
This command deletes a file containing an ACLR standard.
Suffix:
<n>
.
Window
<m> Marker
<sb> irrelevant
Parameters:
<Standard> String containing the file name of the standard.
Manual operation: See " User Standards " on page 160
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:SAVE <Standard>
This command saves the current ACLR measurement configuration as a new ACLR
standard.
The measurement configuration for power measurements consists of weighting filter,
channel bandwidth and spacing, resolution and video bandwidth, detector and sweep
time.
Suffix:
<n>
.
Window
Configuring and Performing Measurements