Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
762User Manual 1178.8520.02 ─ 01
Parameters:
<LowerLimit> The limit of the lower adjacent channel.
Range: 0 dB to 100 dB
*RST: 0 dB
Default unit: dB
<UpperLimit> The limit of the upper adjacent channel.
Range: 0 dB to 100 dB
*RST: 0 dB
Default unit: dB
Manual operation: See " Limit Check " on page 165
See " Limit Checking " on page 177
CALCulate<n>:LIMit<li>:ACPower:ACHannel:RESult?
This command queries the state of the limit check for the adjacent channels in an
ACLR measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also INITiate<n>:CONTinuous on page 746.
Suffix:
<n>
.
irrelevant
<li> irrelevant
Return values:
<LowerACH> text value
Thestate of the limit check for the lower adjacent channels.
PASSED
Limit check has passed.
FAIL
Limit check has failed.
<UpperACH> text value
The state of the limit check for the upper adjacent channels.
PASSED
Limit check has passed.
FAIL
Limit check has failed.
Example:
INIT:IMM;*WAI;
CALC:LIM:ACP:ACH:RES?
PASSED,PASSED
Usage: Query only
Manual operation: See " Limit Check " on page 165
See " Limit Checking " on page 177