Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
763User Manual 1178.8520.02 ─ 01
CALCulate<n>:LIMit<li>:ACPower:ACHannel[:RELative]:STATe
This command turns the relative limit check for the adjacent channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe].
Suffix:
<n>
.
irrelevant
<li> irrelevant
Manual operation: See " Limit Check " on page 165
See " Limit Checking " on page 177
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:ABSolute <LowerLimit>[,
<UpperLimit>]
This command defines the absolute limit of the alternate channels.
If you have defined an absolute limit as well as a relative limit, the R&S FSV/A uses the
lower value for the limit check.
Suffix:
<n>
.
irrelevant
<li> irrelevant
<ch> 1..n
Alternate channel number
Parameters:
<LowerLimit> The limit of the lower adjacent channel.
Range: -200 dBm to 200 dBm
*RST: -200 dBm
Default unit: dBm
<UpperLimit> The limit of the upper adjacent channel.
Range: -200 dBm to 200 dBm
*RST: -200 dBm
Default unit: dBm
Manual operation: See " Limit Check " on page 165
See " Limit Checking " on page 177
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:ABSolute:STATe
This command turns the absolute limit check for the alternate channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe].