Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
798User Manual 1178.8520.02 ─ 01
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Frequency> Numeric value.
*RST: -12.75 MHz (range 1), -2.515 MHz (range 2), 2.515
MHz (range 3)
Default unit: Hz
Manual operation: See " Range Start / Range Stop " on page 222
See " Range Start / Range Stop " on page 256
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STOP
<Frequency>
This command defines the stop frequency of a spurious emission measurement range.
Make sure to set an appropriate span. If you set a span that is
smaller than the span the sweep list covers, the R&S FSV/A will not measure the
ranges that are outside the span - results may be invalid.
greater than the span the sweep list covers, the R&S FSV/A will adjust the start fre-
quency of the first range and the stop frequency of the last range to the span
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Frequency> Numeric value.
*RST: -2.52 MHz (range 1), 2.52 MHz (range 2), 250.0
MHz (range 3)
Default unit: Hz
Manual operation: See " Range Start / Range Stop " on page 222
See " Range Start / Range Stop " on page 256
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:ATTenuation <Attenuation>
This command defines the input attenuation for a SEM range.
In case of high speed measurements, the input attenuation has to be identical for all
ranges.
Suffix:
<sb>
.
1..n
<ri> 1..n
Selects the measurement range.