Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
828User Manual 1178.8520.02 ─ 01
Suffix:
<ri>
.
irrelevant
Parameters:
<State> ON | 1
The R&S FSV/A stops after measuring one range.
To continue with the next range, you have to use
INITiate<n>:CONMeas.
OFF | 0
The R&S FSV/A sweeps all ranges in one go.
*RST: 0
Manual operation: See " Stop After Sweep " on page 258
[SENSe:]LIST:RANGe<ri>:COUNt?
This command queries the number of ranges in the sweep list (<range> is irrelevant).
Suffix:
<ri>
.
Selects the measurement range.
Return values:
<Ranges> Number of ranges in the sweep list.
Usage: Query only
[SENSe:]LIST:RANGe<ri>:DELete
This command removes a range from the sweep list.
Note that
●
you cannot delete the reference range
●
a minimum of three ranges is mandatory.
Suffix:
<ri>
.
1..n
Selects the measurement range.
[SENSe:]LIST:RANGe<ri>:DETector <Detector>
This command selects the detector for a spurious emission measurement range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Detector> APEak
Autopeak
NEGative
minimum peak detector
Configuring and Performing Measurements