Manual operation: See " Center Frequency = Marker Frequency " on page 442
CALCulate<n>:MARKer<m>:FUNCtion:CSTep
This command matches the center frequency step size to the current marker fre-
quency.
The command turns delta markers into normal markers.
Suffix:
<n>
.
Window
<m> Marker
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:X:SPACing <Scale>
This command selects the scaling of the x-axis.
Suffix:
<n>
.
Window
<w> subwindow
<t>
Parameters:
<Scale> LOGarithmic
Logarithmic scaling.
LINear
Linear scaling.
*RST: LINear
Example:
DISP:TRAC:X:SPAC LOG
Manual operation: See " Frequency Axis Scaling " on page 362
[SENSe:]FREQuency:CENTer <Frequency>
This command defines the center frequency.
Parameters:
<Frequency> The allowed range and f
max
is specified in the data sheet.
UP
Increases the center frequency by the step defined using the
[SENSe:]FREQuency:CENTer:STEP command.
DOWN
Decreases the center frequency by the step defined using the
[SENSe:]FREQuency:CENTer:STEP command.
*RST: fmax/2
Default unit: Hz