Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
889User Manual 1178.8520.02 ─ 01
Parameters:
<Span> The minimum span for measurements in the frequency domain
is 10 Hz.
Range: 0 Hz to fmax
*RST: Full span
Default unit: Hz
Manual operation: See " Zero Span " on page 123
See " Span " on page 362
See " Zero Span " on page 363
See " Last Span " on page 363
[SENSe:]FREQuency:SPAN:FULL
This command restores the full span.
Manual operation:
See " Full Span " on page 362
[SENSe:]FREQuency:STARt
<Frequency>
This command defines a start frequency for measurements in the frequency domain.
Parameters:
<Frequency> 0 to (fmax - min span)
*RST: 0
Default unit: HZ
Example:
FREQ:STAR 20MHz
Manual operation: See " Frequency Sweep " on page 122
See " Start / Stop " on page 362
[SENSe:]FREQuency:STOP <Frequency>
This command defines a stop frequency for measurements in the frequency domain.
Parameters:
<Frequency> min span to fmax
*RST: fmax
Default unit: HZ
Example:
FREQ:STOP 2000 MHz
Manual operation: See " Frequency Sweep " on page 122
See " Start / Stop " on page 362
14.7.1.2 Configuring Signal Tracking
When signal tracking is activated, the maximum signal is determined after each fre-
quency sweep and the center frequency is set to the frequency of this signal. Thus with
drifting signals the center frequency follows the signal.