EasyManua.ls Logo

R&S FSV3000 Series - Page 888

R&S FSV3000 Series
1268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
888User Manual 1178.8520.02 ─ 01
RBW
Couples the step size to the resolution bandwidth. Available for
measurements in the time domain.
OFF
Decouples the step size.
*RST: SPAN
Example: //Couple step size to span
FREQ:CENT:STEP:LINK SPAN
Manual operation: See " Center Frequency Stepsize " on page 363
[SENSe:]FREQuency:CENTer:STEP:LINK:FACTor
<Factor>
This command defines a step size factor if the center frequency step size is coupled to
the span or the resolution bandwidth.
Parameters:
<Factor> 1 to 100 PCT
*RST: 10
Default unit: PCT
Example: //Couple frequency step size to span and define a step size fac-
tor
FREQ:CENT:STEP:LINK SPAN
FREQ:CENT:STEP:LINK:FACT 20PCT
Manual operation: See " Center Frequency Stepsize " on page 363
[SENSe:]FREQuency:OFFSet <Offset>
This command defines a frequency offset.
If this value is not 0 Hz, the application assumes that the input signal was frequency
shifted outside the application. All results of type "frequency" will be corrected for this
shift numerically by the application.
See also " Frequency Offset " on page 364.
Parameters:
<Offset> Range: -100 GHz to 100 GHz
*RST: 0 Hz
Default unit: HZ
Example:
FREQ:OFFS 1GHZ
Manual operation: See " Frequency Offset " on page 364
[SENSe:]FREQuency:SPAN <Span>
This command defines the frequency span.
If you set a span of 0 Hz in the Spectrum application, the R&S FSV/A starts a mea-
surement in the time domain.
Setting Basic Measurement Parameters