Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
907User Manual 1178.8520.02 ─ 01
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition <Position>
This command defines the vertical position of the reference level on the display grid
(for all traces).
The R&S FSV/A adjusts the scaling of the y-axis accordingly.
For measurements with the optional external generator control, the command defines
the position of the reference value.
Suffix:
<n>
.
Window
<t> irrelevant
Parameters:
<Position> 0 PCT corresponds to the lower display border, 100% corre-
sponds to the upper display border.
*RST: 100 PCT = frequency display; 50 PCT = time dis-
play
Default unit: PCT
Example:
DISP:TRAC:Y:RPOS 50PCT
Manual operation: See " Reference Position " on page 343
See " Ref Level Position " on page 374
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing <ScalingType>
This command selects the scaling of the y-axis (for all traces, <t> is irrelevant).
Suffix:
<n>
.
Window
<w> subwindow
<t> Trace
Parameters:
<ScalingType> LOGarithmic
Logarithmic scaling.
LINear
Linear scaling in %.
LDB
Linear scaling in the specified unit.
PERCent
Linear scaling in %.
*RST: LOGarithmic
Example:
DISP:TRAC:Y:SPAC LIN
Selects linear scaling in %.
Manual operation: See " Scaling " on page 375
Setting Basic Measurement Parameters