Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
915User Manual 1178.8520.02 ─ 01
ON | 1
Switches the function on
*RST: 0
Example:
SWE:EGAT ON
Activate gating
SWE:EGAT:CONT:STAT ON
Activate continuous gating
Manual operation: See "Continuous Gate" on page 407
[SENSe:]SWEep:EGATe:HOLDoff <DelayTime>
This command defines the delay time between the gate signal and the continuation of
the measurement.
Note: If you perform gated measurements in combination with the IF Power trigger, the
R&S FSV/A ignores the holding time for frequency sweep, FFT sweep, zero span and
I/Q mode measurements.
Parameters:
<DelayTime> Range: 0 s to 30 s
*RST: 0 s
Default unit: S
Example:
SWE:EGAT:HOLD 100us
Manual operation: See " Gate Delay " on page 406
[SENSe:]SWEep:EGATe:LENGth <GateLength>
This command defines the gate length.
Parameters:
<GateLength> Range: 125 ns to 30 s
*RST: 400μs
Default unit: S
Example:
SWE:EGAT:LENG 10ms
Manual operation: See " Gate Length " on page 406
[SENSe:]SWEep:EGATe:POLarity <Polarity>
This command selects the polarity of an external gate signal.
The setting applies both to the edge of an edge-triggered signal and the level of a
level-triggered signal.
Parameters:
<Polarity> POSitive | NEGative
*RST: POSitive
Example:
SWE:EGAT:POL POS
Setting Basic Measurement Parameters