Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
914User Manual 1178.8520.02 ─ 01
Example:
SWE:EGAT ON
Switches on the gate mode.
SWE:EGAT:TYPE EDGE
Switches on the edge-triggered mode.
SWE:EGAT:HOLD 100US
Sets the gate delay to 100 µs.
SWE:EGAT:LEN 500US
Sets the gate opening time to 500 µs.
INIT;*WAI
Starts a sweep and waits for its end.
Manual operation: See " Gated Trigger " on page 406
[SENSe:]SWEep:EGATe:CONTinuous:PCOunt <Amount>
Defines the number of gate periods to be measured after a single trigger event.
Parameters:
<Amount> integer
Range: 1 to 1023
Increment: 1
*RST: 100
Example:
SWE:EGAT:CONT:PCO 50
Manual operation: See "Gate Period Count" on page 407
[SENSe:]SWEep:EGATe:CONTinuous:PLENgth <Time>
Defines the length in seconds of a single gate period in continuous gating. The length
is determined from the beginning of one gate measurement to the beginning of the
next one.
Parameters:
<Time> Range: 125 ns to 30 s
*RST: 5 ms
Default unit: S
Example:
SWE:EGAT:CONT:PLEN 10
Manual operation: See "Gate Period Length" on page 407
[SENSe:]SWEep:EGATe:CONTinuous[:STATe] <State>
Activates or deactivates continuous gating.
This setting is only available if [SENSe:]SWEep:EGATe is "On".
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
Setting Basic Measurement Parameters