Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
913User Manual 1178.8520.02 ─ 01
Parameters:
<Interval> 2.0 ms to 5000
Range: 2 ms to 5000 s
*RST: 1.0 s
Default unit: S
Example:
TRIG:SOUR TIME
Selects the time trigger input for triggering.
TRIG:TIME:RINT 50
The sweep starts every 50 s.
Manual operation: See " Repetition Interval " on page 399
14.7.4.2
Configuring Gated Measurements
[SENSe:]SWEep:EGATe................................................................................................. 913
[SENSe:]SWEep:EGATe:CONTinuous:PCOunt.................................................................914
[SENSe:]SWEep:EGATe:CONTinuous:PLENgth................................................................914
[SENSe:]SWEep:EGATe:CONTinuous[:STATe]..................................................................914
[SENSe:]SWEep:EGATe:HOLDoff....................................................................................915
[SENSe:]SWEep:EGATe:LENGth.....................................................................................915
[SENSe:]SWEep:EGATe:POLarity....................................................................................915
[SENSe:]SWEep:EGATe:SOURce................................................................................... 916
[SENSe:]SWEep:EGATe:TYPE........................................................................................916
[SENSe:]SWEep:EGATe <State>
This command turns gated measurements on and off.
The measurement ends when a particular number of measurement points has been
recorded.
(See [SENSe:]SWEep[:WINDow<n>]:POINts on page 896).
Performing gated measurements turns the squelch off.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Setting Basic Measurement Parameters