Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
933User Manual 1178.8520.02 ─ 01
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<Offset> <numeric value>, specified in Hz, kHz, MHz or GHz, rounded to
the nearest Hz
*RST: 0 Hz
Default unit: HZ
Example: //Define an offset between generator output frequency and ana-
lyzer frequency
SOUR:EXT:FREQ:OFFS 10HZ
Manual operation: See "(Automatic) Source Frequency (Numerator/Denominator/
Offset)" on page 340
SOURce<si>:EXTernal<gen>:POWer[:LEVel]
<Level>
This command sets the output power of the selected generator.
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<Level> <numeric value>
*RST: -20 dBm
Default unit: DBM
Example: //Define generator output level
SOUR:EXT:POW -30dBm
Manual operation: See " Source Power " on page 339
SOURce<si>:EXTernal<gen>[:STATe] <State>
This command activates or deactivates the connected external generator.
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation: See " Source State " on page 339