Common Measurement Settings
R&S
®
FSVA3000/ R&S
®
FSV3000
339User Manual 1178.8520.02 ─ 01
Source State .............................................................................................................. 339
Source Power .............................................................................................................339
Source Offset ............................................................................................................. 339
Source Frequency Coupling........................................................................................340
(Manual) Source Frequency........................................................................................340
(Automatic) Source Frequency (Numerator/Denominator/Offset)...............................340
Result Frequency Start .............................................................................................. 341
Result Frequency Stop ...............................................................................................341
Source State
Activates or deactivates control of an external generator.
Remote command:
SOURce<si>:EXTernal<gen>[:STATe] on page 933
Source Power
The output power of the external generator. The default output power is -20 dBm. The
range is specified in the data sheet.
Remote command:
SOURce<si>:EXTernal<gen>:POWer[:LEVel] on page 933
Source Offset
Constant level offset for the external generator. Values from -200 dB to +200 dB in 1
dB steps are allowed. The default setting is 0 dB. Offsets are indicated by the "LVL"
label in the channel bar (see also "Displayed Information and Errors" on page 335).