EasyManua.ls Logo

R&S FSV3000 Series - Page 943

R&S FSV3000 Series
1268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
943User Manual 1178.8520.02 ─ 01
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe]................................................................ 947
[SENSe:]PMETer<p>:ROFFset[:STATe]............................................................................947
[SENSe:]PMETer<p>[:STATe]..........................................................................................947
[SENSe:]PMETer<p>:UPDate[:STATe]..............................................................................948
UNIT<n>:PMETer<p>:POWer..........................................................................................948
UNIT<n>:PMETer<p>:POWer:RATio................................................................................ 949
CALibration:PMETer<p>:ZERO:AUTO ONCE
This command zeroes the power sensor.
Note that you have to disconnect the signals from the power sensor input before you
start to zero the power sensor. Otherwise, results are invalid.
Suffix:
<p>
.
Power sensor index
Example:
CAL:PMET2:ZERO:AUTO ONCE;*WAI
Starts zeroing the power sensor 2 and delays the execution of
further commands until zeroing is concluded.
Usage: Event
Manual operation: See " Zeroing Power Sensor " on page 319
CALCulate<n>:PMETer<p>:RELative[:MAGNitude] <RefValue>
This command defines the reference value for relative measurements.
Suffix:
<n>
.
Window
<p> Power sensor index
Parameters:
<RefValue> Range: -200 dBm to 200 dBm
*RST: 0
Default unit: DBM
Example:
CALC:PMET2:REL -30
Sets the reference value for relative measurements to -30 dBm
for power sensor 2.
Manual operation: See " Reference Value " on page 321
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE
This command sets the current measurement result as the reference level for relative
measurements.
Suffix:
<n>
.
Window
<p> Power sensor index
Setting Basic Measurement Parameters