Loading...
Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
995User Manual 1178.8520.02 ─ 01
Suffix:
<n>
.
Window
<m> Marker
Manual operation: See " Peak Search " on page 441
CALCulate<n>:MARKer<m>:MAXimum:RIGHt
This command moves a marker to the next lower peak.
The search includes only measurement values to the right of the current marker posi-
tion.
In the spectrogram, the command moves a marker horizontally to the maximum level in
the currently selected frame. The vertical marker position remains the same.
Suffix:
<n>
.
Window
<m> Marker
Manual operation: See " Search Next Peak " on page 441
CALCulate<n>:MARKer<m>:MINimum:AUTO <State>
This command turns an automatic marker peak search for a trace minimum on and off.
The R&S FSV/A performs the peak search after each sweep.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK:MIN:AUTO ON
Activates the automatic minimum value search function for
marker 1 at the end of each particular sweep.
Manual operation: See " Auto Max Peak Search / Auto Min Peak Search "
on page 436
CALCulate<n>:MARKer<m>:MINimum:LEFT
This command moves a marker to the next minimum value.
The search includes only measurement values to the right of the current marker posi-
tion.