Remote Control Commands
R&S
®
SMB100A
366Operating Manual 1407.0806.32 ─ 21
Parameters:
<Frequency> <Frequency#1>{, <Frequency#2>, ...} | block data
The data can be given either as a list of numbers (list can be of
any length and list entries must be separated by commas) or as
binary block data. When block data is transferred, 8 bytes are
always interpreted as a floating-point number with double accu-
racy (see :FORMat[:DATA] on page 294).
Range: 300 kHz to RFmax
Example:
LIST:SEL '/var/list3'
selects list3 for editing. The R&S SMB generates a new file
automatically, if it does not exist yet.
SOUR:LIST:FREQ 1.4GHz, 1.3GHz, 1.2GHz,...
specifies the frequency values in list3. If the list already contains
data, it is overwritten.
Manual operation: See "Edit List Mode Data... - List Mode" on page 192
[:SOURce<hw>]:LIST:FREQuency:POINts?
The command queries the length (in points) of the FREQuency component of the
selected list.
Return values:
<Points> integer
Range: 0 to INT_MAX
*RST: 0
Example:
LIST:SEL '/var/list3'
selects list3 for editing. The R&S SMB creates a new file
automatically, if it does not exist yet.
LIST:FREQ:POIN?
queries the number of frequency values in the list
Response: 327
Usage: Query only
[:SOURce<hw>]:LIST:INDex <Index>
Sets the list index in step mode (LIST:MODE STEP).
After the trigger signal the frequency and level settings of the selected index are pro-
cessed in List mode.
Parameters:
<Index> integer
*RST: 0
SOURce Subsystem