Fernsteuerbefehle
R&S
®
RTB2000
290Bedienhandbuch 1333.1611.03 ─ 05
Parameter:
<AutomaticRoll> ON | OFF
*RST: OFF
TIMebase:ROLL:MTIMe <MinTimeBase>
The roll mode is enabled automatically if the time base exceeds the MinTimeBase,
and if TIMebase:ROLL:AUTomatic auf Seite 289 is set ON.
Parameter:
<MinTimeBase> Limit value for roll mode enabling.
*RST: 500e-3
Std.-einheit: s/div
ACQuire:INTerpolate <InterpolationType>
Defines the interpolation mode.
See also: "Interpolation" auf Seite 55
Parameter:
<InterpolationType> SINX | LINear | SMHD
LINear
Linear interpolation between two adjacent sample points.
SINX
Interpolation by means of a sin(x)/x curve.
SMHD
Sample & hold causes a histogram-like interpolation.
*RST: SINX
ACQuire:POINts:ARATe?
Retrieves the sample rate of the ADC, that is the number of points that are sampled by
the ADC in one second.
Rückgabewerte:
<ACDsampleRate> ADC sample rate
Std.-einheit: Hz
Verwendung: Nur Abfrage
ACQuire:SRATe? <SampleRate>
Returns the sample rate, that is the number of recorded waveform samples per
second.
Parameter:
<SampleRate> Std.-einheit: Sa/s
Verwendung: Nur Abfrage