Fernsteuerbefehle
R&S
®
RTB2000
442Bedienhandbuch 1333.1611.03 ─ 05
Suffix: 
<b>
.
1 | 2
<n> *
Selects the frame.
<o> *
Selects the word number.
Rückgabewerte: 
<Status> OK | FRSTart | FRENd | FRMError | STERror | SPERror |
PRERror | INSufficient | BREak
OK: the frame is valid.
FRSTart: frame start not found
FRENd: frame end not found
FRMError: error in frame
STERror: start error, no start bit found.
SPERror: stop error, no stop condition found.
PRERror: parity error, which indicates a transmission error.
INSufficient: the frame is not completely contained in the
acquisition. The acquired part of the frame is valid.
BREak: break condition found. A start bit is not followed by a
stop bit, and the data line remains at logic 0 for longer than a
UART word.
Verwendung: 
Nur Abfrage
BUS<b>:UART:FRAMe<n>:WORD<o>:STARt?
BUS<b>:UART:RX:FRAMe<n>:WORD<o>:STARt?
BUS<b>:UART:TX:FRAMe<n>:WORD<o>:STARt?
Returns the start time of the specified symbol (word).
Suffix: 
<b>
.
1 | 2
<n> *
Selects the frame.
<o> *
Selects the word number.
Rückgabewerte: 
<StartTime> Bereich:  depends on sample rate, record length, and time
base
Inkrement:  depends on the time base
Std.-einheit: s
Verwendung:  Nur Abfrage
Analyse serieller Busse