Fernsteuerbefehle
R&S
®
RTB2000
446Bedienhandbuch 1333.1611.03 ─ 05
Parameter:
<TriggerType> STOFrame | EOFrame | ID | IDDT | FTYPe | ERRCondition
STOFrame
Start of frame
EOFrame
End of frame
ID
Sets the trigger to a specific message „Identifier“ or an „Identi-
fier“ range.
Specify the identifier with TRIGger:A:CAN:ITYPe, TRIGger:
A:CAN:ICONdition and TRIGger:A:CAN:IDENtifier.
IDDT
Sets the trigger to a combination of „Adresse und Daten“ (Identi-
fier and Data) condition. The instrument triggers at the end of
the last byte of the specified data pattern.
Specify the „Identifier“ (see ID), and the „Daten“ (Data) with
TRIGger:A:CAN:DLC, TRIGger:A:CAN:DCONdition and
TRIGger:A:CAN:DATA.
FTYPe
Triggers on a specified „Frame“.
Specify the frame type with TRIGger:A:CAN:FTYPe.
ERRCondition
Identifies various errors in the frame.
Specify the „Fehler“ (Error) with TRIGger:A:CAN:ACKerror,
TRIGger:A:CAN:BITSterror, TRIGger:A:CAN:CRCerror
and TRIGger:A:CAN:FORMerror.
*RST:
STOF
TRIGger:A:CAN:FTYPe <FrameType>
Specifies the frame type to be triggered on if TRIGger:A:CAN:TYPE is set to FTYPe.
Parameter:
<FrameType> DATA | REMote | ERRor | OVERload | ANY
*RST: ERR
TRIGger:A:CAN:ITYPe <IdentifierType>
Selects the length of the identifier: 11 bit for CAN base frames, or 29 bits for CAN
extended frames.
The command is relevant if TRIGger:A:CAN:TYPE is set to ID, IDDT, or FTYPe (data
and remote frames).