EasyManua.ls Logo

Rohde & Schwarz RTB2004 - Page 495

Rohde & Schwarz RTB2004
547 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fernsteuerbefehle
R&S
®
RTB2000
495Bedienhandbuch 1333.1611.03 ─ 05
PGENerator:PATTern:COUNter:FREQuency <Period>
Defines the frequency value of the pattern generator counter function. The user fre-
quency always refers to the switching of the pattern condition. This results in square
waveforms for individual pins.
Pin Frequency
S0 f/2
S1 f/4
S2 f/8
S3 f/16
Parameter:
<Period> Numeric value
Bereich: 2.380952425301E-02 bis 2.500000000000E+07
*RST: 1.000000000000E+05
Std.-einheit: Hz
PGENerator:PATTern:COUNter:DIRection <CountDirection>
Sets the pattern generator counter direction.
Parameter:
<CountDirection> UPWard | DOWNward
*RST: UPW
PGENerator:PATTern:SQUarewave:POLarity <Polarity>
Defines the polarity of the pattern generator square wave function.
Parameter:
<Polarity> NORMal | INVerted
*RST: NORM
PGENerator:PATTern:SQUarewave:DCYCle <DutyCycle>
Sets the duty cycle of the square wave function.
Parameter:
<DutyCycle> Numeric value
Bereich: 1.00E+00 bis 9.900E+01
*RST: 5.000E+01
Std.-einheit: %
Beispiel: PGEN:PATT:SQU:DCYC 20
Sets the duty cycle of the square wave function to 20%.
Signalerzeugung (Option R&S RTB-B6)

Table of Contents

Related product manuals