Note:
These Quick Operating Instructions represent only an
overview of the operating procedures. For a complete setup and
use of the machine, please refer to the Xtra Operator’s Manual.
XTRA Operator’s Manual 20077/021 US C-5
2. Connect the Suction Line:
j. Using aseptic technique, spike the container of the anticoagulant solution
3. Set up the Vacuum Line:
a. Attach one end of the vacuum line to the X
VAC system (or another vacuum
source)
3. Set up the Vacuum Line:
b. The other to the vacuum port (yellow capped) on the lid of the reservoir
3. Set up the Vacuum Line:
c. Verify that the protective cover on the end of the suction line in the sterile
field has been removed and that the aspiration tip is open to the
atmosphere (not blocked).
3. Set up the Vacuum Line:
d. With the X
VAC system ON, open the roller clamp on the small bore tubing
and allow 200 ml of anticoagulant solution to be drawn into the Blood
Collection Reservoir. This volume should ensure adequate wetting of the
blood contact surfaces. Close the roller clamp if there will be any delay
before processing.
4. Install the Bowl Set:
a. Hang the tray onto the two handles placed on the top right panel of the
machine and open it
4. Install the Bowl Set:
b. Open the centrifuge lid
4. Install the Bowl Set:
c. Swing open the centrifuge arm
4. Install the Bowl Set:
d. Unlatch and open the clamp lid
4. Install the Bowl Set:
e. Remove the bowl/tubing harness from the tray. Remove the protective
spacer from the upper part of the bowl
Setting Up the Disposables
Quick Start Instructions: Setting Up the Disposables