Chapter 12 — Library Troubleshooting Resetting Components
December 2019 User Guide—Spectra T200, T380, and T680 Libraries
360
Reset a RIM or F-QIP
The reset procedures for a RIM and an F-QIP are identical.
Hardware Reset
To perform a hardware reset of a RIM or F-QIP, insert the tip of a toothpick
or other similar non-conductive object into the hole in the back panel to
press the Reset button.
BlueScale Software Reset
You can also reset the RIM or F-QIP through the BlueScale user interface.
1. Determine the BlueScale identifier for the controller you need to reset.
Note: RIMs are shown as F-QIPs in the BlueScale user interface and are
identified according to their physical location in the library. See
RIM and F-QIP Identifiers on page 57 for information about
identifiers.
2. Log into the library as a user with superuser or administrator
privileges.
3. From the toolbar menu, select Maintenance > Tools. The Maintenance
Tools screen displays.
4. Select Utilities. The Utilities screen displays.
Caution
Do not use a metallic object to press the reset button. You may inadvertently cause
an electrical short inside the module.
Figure 214 Location of the RIM or F-QIP hardware reset button.
Reset
RIM
Reset
F-QIP
Reset
RIM2