Chapter 12 — Library Troubleshooting Discovering and Setting the BlueScale Vision Camera IP Address
December 2019 User GuideSpectra T200, T380, and T680 Libraries
373
2. Locate the recessed Reset on the camera.
Notes: If the transporter is blocking access to the camera, carefully
move it out of the way by gently pushing on the VAX column
near where the transporter is mounted.
To simplify identifying the camera when using the camera
setup wizard, make a note of the last six digits on the camera’s
MAC address. A label with the MAC address is located near
the camera’s Ethernet connector. Select Administrator >
Status to view the MAC address through the Network
Camera Setup wizard (see Discovering and Setting the
BlueScale Vision Camera IP Address on page 362).
3. Insert the tip of a toothpick or other similar non-conductive object into
the hole to press the reset button. Hold the reset button for 10 seconds
to reset the camera's settings to the factory defaults, which include
using DHCP to obtain an IP address.
4. Replace the library’s side panel as described in Install the Side Access
Panel on page 388.
Figure 228 The BlueScale Vision camera Reset button (viewed from left side of
library).
Reset button
MAC address label
Caution
Do not use a metallic object to press the reset button. You may inadvertently cause
an electrical short inside the camera.