3.3 Timer
HA-8190V 3-5
c Set the startup and shutdown timers.
• See the table below.
d Touch .
• The message “Settings changed. Save setting changes?” will
appear.
e Touch .
• This saves your new settings and return you to the [Set timer]
screen.
f Touch to return to the standby screen.
Go to the section below that works with your <Startup timer> setting.
<Use>: “ When using the startup timer” on page 3-6.
<Not use>: “ When NOT using the startup timer” on page 3-7
REFERENCE:
To go to page 2/2:
Touch [].
OK
Cancel
Setup item Description (default bolded)
Startup timer Use: Activates the startup timer.
Not use: Does not activate the startup timer.
Monday to Sunday Start up /
Not start up
When <Use> is selected for <Startup timer>, select whether to start up the
instrument or not on each day of the week.
Start up: Automatically starts up the instrument.
Not start up: Does not start up the instrument automatically.
Start time When <Start up> is selected for <Start up / Not start up>, set what time to
start up the instrument on each day of the week.
Range of time: 00:00 to 23:59
Shutdown timer
(Time until auto-sleep)
Set the amount of time until the instrument sleeps when no operations are performed with
the standby screen displayed. Set to “00:00” if you do not use the shutdown timer. When
the instrument sleeps due to the shutdown timer, it will start up on the first available day set
as <Start up>.
Range: 00:00 (Not use) to 23:59