Chapter 3 Auxiliary Operations
HA-8190V 3-12
[Diagnosis menu] screen (8000)
[Meas. counter setup] screen (9000)
Option Description (default bolded) Page
Tube test menu
(8100)
All (8110): Tests the drive unit, sample introduction flow and
drain flow once each.
3-33
Drive unit test (8120): Tests the drive unit. 3-34
Sample intro flow test (8130): Tests the sample introduction flow.
Drain flow test (8140): Tests the drain flow.
Accuracy control
(8200)
Prints out statistical information on HbA1c control measurements and sample
measurements.
3-35
Date (DD-MM-YY): Range of measurement dates (01-01-2000 to 31-12-2099)
Meas. mode: Fast, Variant (current measurement mode)
Analysis section
check (8300)
Checks that the analysis section is working properly. 3-36
Reproducibility menu
(8400)
Whole sample meas.(8410):
Repeatedly measures a particular whole blood sample and displays statistical
information.
3-37
Hemolysis sample meas.(8420):
Measures the same hemolysis sample or hemolysis control divided into multiple
sample containers and displays statistical information.
3-39
Description Page
Sets the column counter to a desired number. 3-50