Chapter 4 Maintenance
HA-8190V 4-4
a Enter the 10-digit reagent information code that is written on the
label of a new eluent pack.
If your entry is accepted, the lot number and expiration date will
be updated.
3 Remove the used eluent pack.
a Lay out some gauze near the instrument.
b Remove the used eluent pack from the bottle tray.
c Remove the bottle cap with nozzle from the pack.
Place the nozzle on the gauze.
4 Place the new eluent pack.
a Hold the new eluent pack by the hard plastic neck and remove the
cap from the pack.
b Wipe any liquid from the nozzle with a new piece of gauze.
Reagent code on
the pack label
NOTE:
If <Lot No.> and <Expiry> are not updated and still contain
hyphens (---):
Your entry may be incorrect. Carefully check the reagent
code and enter it again.
REFERENCE:
You can use the optional hand-held barcode reader to read the
reagent information code.
Hold here.
NOTE:
Do not hold the eluent pack by the soft aluminium bag.
Eluent may spill and damage the instrument.
REFERENCE:
Keep the cap in the accessory case. This cap should be reused when transporting the instrument, or if the
instrument is not to be used for extended periods of time.
NOTE:
Remove any lint if it is stuck to the nozzle. Lint may clog the tube.