EasyManua.ls Logo

Arkray ADAMS A1c - Page 180

Arkray ADAMS A1c
288 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.2 Replacement of Consumables
HA-8190V 4-15
8 Check for liquid leaks.
a Check that liquid does not leak from the connections on the IN and OUT sides of the column.
Retighten the push screws if liquid leaks.
9 Close the covers.
a Close the temperature control box cover, then the front cover.
10 Complete maintenance.
a Touch .
b When “Has the column been replaced?” appears, touch
.
The column counter will be reset.
The instrument will record the maintenance date.
Mechanical initialization will start.
When initialization is complete, the message “Fluid runs at the
same rate as during measurement.” will appear.
When fluid pumping is complete, the [Reagent replace menu]
screen will appear again.
Finish
REFERENCE:
The message “W-062 The front or maintenance cover is
open.” appears if is touched with the front cover
open. Be sure to close the front cover before touching
.
Touch if you do not reset the column counter.
OK
OK
OK
Cancel

Table of Contents

Related product manuals