Chapter 4 Maintenance
HA-8190V 4-16
c Touch to return to the standby screen.
Check that [Column counter] is reset to “0000”.
d Perform HbA1c calibration.
See “2.8. HbA1c Calibration” on page 2-38.
REFERENCE:
The [Column counter] value will not be reset if was touched in step 10-b.
HOME
Cancel