120
Function
Prompt
Selections or
Range of Setting
Parameter Definition
ATTENTION: For Burnout to function
properly on a 0-20 mA input type (or a 0-
5V type that uses a dropping resistor), the
dropping resistor must be remotely
located (across the transmitter terminals).
Otherwise, the input at the instrument
terminals will always be 0 mA (i.e., within
the normal operating range) when the 0-
20 mA line is opened.
ATTENTION: The Thermocouple Health
feature is disabled when No Failsafe is
configured.
Chapter 4 - Configuration